In vitro study of Magnetic Resonance Imaging artifacts of anesthetic devices

Authors

  • Uday Bajracharya Shree Birendra Hospital, Chhauni and National Academy of Medical Sciences, Bir Hospital
  • Prabhat Rawal Nepal Armed Police Force Hospital, Balambu, Kathmandu

Keywords:

anesthesia, artifacts, Magnetic Resonance Imaging

Abstract

Background: Deep sedation or general anesthesia is usually required for Magnetic Resonance Imaging when patients cannot remain motionless in the suite. Various anesthetic devices have been used to maintain the airway and ventilate the lungs during this period. Some of them produce artifacts that pose difficulties in the interpretation of images. The aim of this study was to identify the devices that produced artifacts during Magnetic Resonance Imaging.

Methods: Twelve anesthetic devices were considered: oro-pharyngeal airway, naso-pharygeal airway, face mask with reservoir bag, nasal cannula, endotracheal tube, disposable Ambu Laryngeal Mask Airway, Laryngeal Mask Airway Unique, Disposable Laryngeal Tube Sonda, i-gel, Ambu bag, Bain Circuit, Jackson Rees Circuit. Magnetic Resonance Imaging was performed with each device placed on the top of a phantom simulator respectively to resemble the position in vivo.

Results: The artifacts with Disposable Laryngeal Tube Sonda, Laryngeal Mask Airway Unique and endotracheal tube were related to ferromagnetic material in the pilot valve and were similar. No artifacts were found with oro-pharyngeal airway, naso-pharygeal airway, nasal cannula, endo-tracheal tube with pilot valve detached, face masks with reservoir bag (metal removed), Ambu bag (without Adjustable Pressure Limiting valve), i-gel, disposable Ambu Laryngeal Mask Airway, Bain Circuit and Jackson Rees Circuit.

Conclusion: Anesthetic devices that produce artifacts Magnetic Resonance Imaging are Disposable Laryngeal Tube Sonda, Laryngeal Mask Airway Unique and Endotracheal Tube.

Author Biographies

Uday Bajracharya, Shree Birendra Hospital, Chhauni and National Academy of Medical Sciences, Bir Hospital

Senior Consultant Anesthesiologist,

Department of Anesthesiology,

Shree Birendra Hospital, Chhauni, Kathmandu.

Prabhat Rawal, Nepal Armed Police Force Hospital, Balambu, Kathmandu

Anesthesiologist, Nepal APF Hospital.

References

1. Gooden CK, Dilos B. Anesthesia for magnetic resonance imaging. Int Anesthesiol Clin 2003;41:29-37.
2. Boutros A, Pavlicek W. Anesthesia for magnetic resonance imaging. Anesth Analg 1987;66:367.
3. Osborn IP. Magnetic resonance imaging anesthesia: new challenges and techniques. Curr Opin Anaesthesiol 2002;15:443-8.
4. Crofts S, Campbell A. A source of artefact during general anaesthesia for magnetic resonance imaging. Anaesthesia 1993;48:643.
5. Fairfield JE. Laryngeal Mask and magnetic resonance- a caution. Anesthesia 1990;45:995.
6. Langton JA, Wilson I, Fell D. Use of the laryngeal mask airway during magnetic resonance imaging. Anesthesia 1992;47:532.
7. Anez C, Fuentes A, Jubera P, Sala JM, Tull M. The proseal laryngeal mask airway interferes with magnetic resonance imaging. Can J Anaesth 2005;52:116-7.
8. Stevens JE. Burden G. Reinforced laryngeal mask airway and magnetic resonance imaging. Anesthesia 1994;49:79-80.
9. Monclus E, Garces A, De Jose Maria B, Artes D, Mabrock M. Study of the adjustment of the Ambu laryngeal mask under magnetic resonance imaging. Pediatr Anesth 2007;17:1182-6.
10. Taxak S, Bhardwaj M, Gopinath A. The i-gel - A promising airway device for magnetic resonance imaging suite. J Anaesthesiol Clin Pharmacol 2012;28:263-4.
11. Schieble T, Patel A, Davidson M. Laryngeal mask airway (LMA) artefact resulting in MRI misdiagnosis. Pediatr Radiol 2008;38:328-30.
12. Brimacombe JR. Intubating LMA for airway intubation. In: Brimacombe JR, editor. Laryngeal Mask Anesthesia: Principles and practice. 2nd ed. Philadelphia: Saunders; 2005. pp. 469-505.
13. Nixon C, Hirsch NP, Ormerod IEC, Johnson G. Nuclear Magnetic resonance; its implications for the anesthetist. Anaesthesia 1986;41:131-7.
14. Grady RE Jr, Perkins WJ. An unexpected cause of magnetic resonance image distortion: the endotracheal tube pilot balloon. Anesthesiology 1997;86:993-4.

Downloads

Published

2015-02-26

How to Cite

Bajracharya, U., & Rawal, P. (2015). In vitro study of Magnetic Resonance Imaging artifacts of anesthetic devices. Journal of Society of Anesthesiologists of Nepal, 2(1), 13-16. Retrieved from http://www.jsan.org.np/jsan/index.php/jsan/article/view/31

Issue

Section

Original Article