Damage to endotracheal cuff inflation tube by harmonic scalpel in tonsillectomy leading to compromised airway

Authors

  • Asfar Azimee Shija Hospitals and Research Institute, Meitei Langol, Lamphelpat, Imphal, Manipur 795004
  • Dinesh Singh Thounaojam Shija Hospitals and Research Institute, Meitei Langol, Lamphelpat, Imphal, Manipur 795004

Keywords:

airway management, anaesthesia, complications, endotracheal tube

Abstract

Use of harmonic scalpel has increased in the recent past for different surgical procedures. We report a case of damage to an endotracheal tube inflation line in a 47 year old female patient who underwent elective tonsillectomy due to harmonic scalpel. Pre-anaesthetic evaluation of the patient revealed no abnormalities. The patients trachea was intubated via nasal route with a cuffed north facing Ring, Adair and Elwyn tube of 7.0 mm internal diameter. The cuff was inflated after confirmation of tracheal placement of tube and pharyngeal packing was done. After 30 minutes of surgery, the bellows of the ventilator were collapsing. Suspecting a leak an attempt was made to inflate the cuff, which revealed high resistance in inflation tube. Surgeon was immediately asked to stop the surgery. Check laryngoscopy revealed damage to cuff inflation tube which was sealed by harmonic scalpel in semi-inflated condition. Anticipating that the tracheal extubation in that inflated position would result in vocal cord trauma, a 23 gauge spinal needle attached to 5ml syringe was introduced under direct laryngoscopic guidance. The cuff was punctured and air aspirated so as to deflate the cuff. The tube was taken out and was replaced. The surgery continued for next one hour, was uneventful and trachea extubated at completion.

Author Biographies

Asfar Azimee, Shija Hospitals and Research Institute, Meitei Langol, Lamphelpat, Imphal, Manipur 795004

DA DNB, Attending consultant, Department of Anaesthesiology and Critical Care

Dinesh Singh Thounaojam, Shija Hospitals and Research Institute, Meitei Langol, Lamphelpat, Imphal, Manipur 795004

MD, Senior Consultant, Department of Anaesthesiology & Critical Care

References

1. Rolbin S. An unusual cause of ventilator leak. Can Anaesth Soc J 1977;24:522-4.
http://dx.doi.org/10.1007/BF03005457
[PMID:268230]
2. Lee O, Sommer RM. Pressure monitoring hose causes leak in anaesthesia breathing circuit. Anesth Analg 1991;73:365.
http://dx.doi.org/10.1213/00000539-199109000-00035
[PMID:1867440]
3. Colavita RD, Apfelbaum JL. An unusual source of leak in the anaesthesia circuit. Anesthesiology 1985;62:208-9.
http://dx.doi.org/10.1097/00000542-198502000-00031
4. Cohen D, Dor M. Morbidity and mortality of post-tonsillectomy bleeding: analysis of cases. J Laryngol Otol 2008;122:88-92.
http://dx.doi.org/10.1017/S0022215107006895
[PMID:17349099]
5. Kearl RA, Hooper RG. Massive airway leaks: an analysis of the role of endotracheal tubes. Crit Care Med 1993;21:518-21.
http://dx.doi.org/10.1097/00003246-199304000-00010
[PMID: 8472570]
6. Henderson MA. Airway obstruction with a cuffed single use plastic endotracheal tube. Anaesth Intens Care 1993;21:370-2.
[PMID:8342777]
7. Lewer BMF, Karim Z, Henderson RS. Large air leak from an endotracheal tube due to a manufacturing defect. Anesth Analg 1997;85:944-5.
http://dx.doi.org/10.1213/00000539-199710000-00043
[PMID:9322486]
8. Gettelman TA. Endotracheal tube failure: undetected by routine testing. Anaesth Analg 1995;81:1311-20.
http://dx.doi.org/10.1097/00000539-199512000-00042
[PMID: 7486131]
9. Kee WD. An unusual problem with an endotracheal tube. Anaesth Intens Care 1993;21:247-8.
[PMID: 8517523]

Downloads

Published

2016-03-20

How to Cite

Azimee, A., & Thounaojam, D. S. (2016). Damage to endotracheal cuff inflation tube by harmonic scalpel in tonsillectomy leading to compromised airway. Journal of Society of Anesthesiologists of Nepal, 3(1), 41-43. Retrieved from http://www.jsan.org.np/jsan/index.php/jsan/article/view/86

Issue

Section

Case Report